This document details the procedure for the Operation and Calibration of the UV-VIS Spectrophotometer (PERKIN ELEMER LAMBDA 25).Â
It is the policy of pharmaceuticals that the written procedure shall be followed for the operation and calibration of UV-VIS Spectrophotometer and its use monitored to ensure that the instrument gives consistently accurate and reproducible results in compliance with regulatory requirements.
This
procedure is to be applied at the time of the Operation and Calibration of the UV-VIS spectrophotometer.
RESPONSIBILITY & ACCOUNTABILITY
Persons along with their responsibilities & account-abilities are given below:
01 Executive - Corporate Quality Assurance, To prepare SOPS
02 Trainee
Analyst/ Jr. Analyst/ Analyst/ Sr. Analyst, To
follow the SOP accordingly
03 Manager
– Quality Assurance, To
ensure implementation of SOP.
04 Manager -Quality
5.0 GENERAL INSTRUCTIONS
5.1 PRELIMINARY CHECK
5.1.1
Check and ensure that the instrument is clean and suitable for starting the
operation. Â If it is not, clean and
ensures suitability.Â
5.1.2 The instrument should be located at the ambient temperature. Â
5.1.3
Ensure the instrument and printer are connected to an uninterrupted power supply.
5.1.4
Check the calibration status of the instrument before use.
5.1.5
Report any discrepancy observed during the calibration of the instrument to the Section
Incharge or his representative for corrective and preventive action.
5.1.6
Section In-charge or his representative will report the same to the Manager –
Quality Assurance and notify the defect to service engineer to rectify the
defect.Â
5.1.7
Affix the ‘Under Maintenance’ label on the instrument.
5.2 CLEANING OF CUVETTE
5.2.1
Never use a brush to clean the inside of the cuvette.
5.2.2
Clean the Cuvette using Purified water or any organic solvent (e.g. Acetone or
Methanol) & dry the cuvette
5.2.3
Add about 1 ml of the solution to be measured. Tilt and turn the cuvette so that
the solution has contact with all the surfaces. Discard the solution and repeat
this rinse once more.
5.2.4
Fill the cuvette about ¾ full of the solution you wish to test.
5.2.5
Wipe the outside of the cuvette with a lint-free, soft tissue to remove any
moisture or fingerprints from the outside surface.Â
5.3 PRECAUTIONS:
5.3.1
Do not stare into the path of ultraviolet rays path.
5.3.2
Do not touch the fine surfaces, always use the rough surface of the cells to hold.
5.3.3
Ensure cleanliness of the cells before and after use.Â
5.3.4
Use Purified water to rinse the cells.
5.3.5
Do not use directly Purified water to rinse cells after use of solutions
prepared with immiscible solvents, use ethanol or methanol first
and then water.
5.3.6
Use a lint-free cloth duster for wiping the cells.
5.3.7
Use keeping cells in one direction in the cell holder throughout the test to minimize
the error in the blank and sample solutions.
5.3.8
Do not shake the solutions of samples immediately before measurement.
5.3.9
Ensure the lid of the UV compartment is closed properly.
5.3.10
The temperature of all solutions/reagents used in the test should not differ by
more than 0.5 ºC.
5.3.11
The solvent in the reference cell should be of the same batch as that used to prepare
the solutions.
5.4Â OPERATION PROCEDURE
5.4.1
Before starting check power supply is on.
5.4.2
Switch on the Instrument by using the On/Off present on the rear side of the UV-VIS
Spectrophotometer.
5.4.3
Allow the UV lamp to glow for mins.
5.4.4
Start Perkin Elmer UV by double-clicking the Icon Perkin Elmer UV Win
Lab software.
5.4.5
Insert User name and Password.Â
5.4.6
Click Ok
5.4.7
Each time you start the instrument application (by double-clicking the instrument
icon from the main window) an Instrument option will start.
5.4.8
Click on the Instrument option it will display Lambda 25 folder
5.4.9
Click on Lambda 25 folder it will display Manual Control
5.4.10
Click Manual Control it will display Settings
5.4.11
Click Settings and set Wavelength as per requirement.
5.4.12
Click Optional Mode and set Absorbance or Transmittance as per requirement.
5.4.13
Enter the Wavelength at which you want to perform the Auto zero and then Click Ok. The instrument goes to the selected wavelength.
5.4.14
From Actions select Auto zero Â
5.4.15
Fill the cuvettes with blank / Solvent solution, wipe with a lint-free cloth
duster and place it in both the cell holders & press Auto Zero the current
wavelength will automatically be set to `0’ Abs. (100% T)
5.4.16
then removes the front side cuvette, and discards the blank solution. Rinse the cuvette
with the sample solution. Then fill the cuvette with sample solution & replace it in the cell holder. The monitor will display the Absorbance of the sample
solution at the wavelength displayed.
5.5. Â SCAN TASK- DATA COLLECTION
5.5.1
The Data collection page allows you to define settings such as scan range,
ordinate mode, and slit widthÂ
5.5.2Â for a scan first click the icon Methods (by clicking the software)
5.5.3
Click New, open the file
5.5.4
Select New Method
5.5.5
Enter the name of the method and required text in the description field.
5.5.6
Click Scan press the key next again then click Save & finish press
OK.Â
5.5.7
Click Data collection
5.5.8Â Click
Scan setting
5.5.9
Following Scan settings are available on the Data collection pageÂ
5.5.9.1
Start
5.5.9.2
End
5.5.9.3
Ordinate Mode
5.5.9.4
Slit Width
5.5.9.5
Scan speed
5.5.9.6Â Â Data Interval
5.5.9.7Â Â Number of cycles
5.5.9.8Â Â Cycle as fast as possible
5.5.9.9Â Â Cycle time
5.5.9.10
UV Lamp on
5.5.9.11
UV Lamp off
5.5.9.12
Lamp change at (nm)
5.5.10
Enter sample information such as sample ID, and description as per requirement by
clicking sample information in the data collection file
5.5.11
Click Auto zero by placing blank in the cell holder
5.5.12
Press Start a plot of wavelength vs. absorbance will be displayed
5.5.13
Save the scan method by pressing save the result.
5.5.14
Print this as per your requirement
5.5.15
before Exit press files this will display option Exit. Press exitÂ
5.6 Maintain
Logbook of UV-VIS Spectrophotometer on APL/CQA/SOP-042/FR-01 LogbookÂ
for UV-VIS Spectrophotometer (Appendix I)
6.0 CALIBRATION PROCEDURE
6.1 CALIBRATION
FREQUENCY:Â HALF-YEARLY
6.2Â
CONTROL OF WAVELENGTHS:Â Â Â Â Â Â Â Â Â Â Â
6.2.1
Dilute 11.5 ml of Perchloric acid (70 % w/w) to 100 ml with Purified water (1.4
M HClO4)
6.2.2
Dissolve 0.4 g of Holmium oxide in 5 ml of 1.4 M Perchloric acid with the aid
of heating on a water bath, cool, and dilute to 10 ml with the same
solvent. (4% w/v Holmium Perchlorate solution).
6.2.3Â Alternatively, a ready-made solution can also be used.
6.2.4
Scan the solution in the range of 200 to 700 nm using 1.4 M Perchloric acid as
a blank. Note down the maxima observed in the Calibration Data Sheet.Â
Check the maxima observed against the acceptance criteria.
6.2.5
Acceptance criteria:
Wavelength      Maximum ToleranceÂ
241.15
nm          (240.15 to 242.15 nm)
287.15
nm          (286.15 to 288.15 nm)
361.50
nm          (360.50 to 362 .50 nm)
536.
30 nm         (535.30 to 537.30 nm)
6.2.6
Record the calibration data on the APL/CQA/SOP-042/FR-02 Calibration datasheet. Â (Appendix II)
6.3 CONTROL OF ABSORBANCE
6.3.1
Control of absorbance using potassium dichromate solution :
6.3.2
Procedure: Add 5.4 ml of sulphuric acid in 500 ml of Purified water and dilute to
1000 ml with Purified water (Solution A).
6.3.3Â Further
dilute 100 ml to 2000 ml with Purified water to give a 0.005 M sulphuric acid
solution. (Blank) [Alternatively, a ready-made solution, can also be
used.]
6.3.4Â Dry
Potassium dichromate (A.R. grade) at 130°C to constant weight and weigh a
quantity accurately between 57.0 mg and 63.0 mg into a 1000.0 ml volumetric
flask. Add 0.005 M Sulphuric acid to dissolve and dilute to the mark. (Solution
B)
6.3.5
Also weigh 57.0-63.0 mg of Potassium dichromate UV into 100.0ml
volumetric flask. Add 0.005M sulphuric acid to dissolve and dilute to the
mark (Solution C)
6.3.6Â Check
the absorbance of solution B at the following wavelengths (nm): 235, 257, 313, 350.
6.3.7Â Check
the absorbance of solution C at wavelength 430nm.
6.3.8
Record the absorbance at each wavelength on APL/CQA/SOP-042/FR-02 Calibration
Data sheet (Appendix II).
6.3.9Â Calculate
the value AÂ (1%, 1 cm) as per the following formula :
 A (1 %, 1 cm)   =        Absorbance x  100  x 100/Weight of K2Cr2O7 in mg
6.3.10
Check the value of A at each wavelength against the acceptance criteria given
below :
6.3.11
Acceptance criteria:
Wavelength
 (nm)    Specific absorbance A 1 %
1 cm      Maximum tolerance for absorbance
235Â Â Â Â Â Â Â Â 124.5Â Â Â Â 122.9Â
 to  126.2   EP & IP
257        144.0    142.4  to  145.7   I P
               144.5    142.8  to  146.2   EP
313         48.6     47.0  to   50.3 Â
 IP & EP
350Â Â Â Â Â Â Â Â 106.6Â Â Â Â 104.9Â
 to  108.2   IP
               107.3    105.6  to  109.0   EP
430Â Â Â Â Â Â Â Â 15.9Â Â Â Â Â Â Â 15.7
to 16.1Â Â IP &EP
6.4 LIMIT OF STRAY LIGHT:
6.4.1
Note the absorbance of a 1.2 % w/v solution of Potassium chloride, at a path
length of 1 cm, the absorbance increases steeply between 220 nm and 200 nm and
is greater than 2 at a wavelength between 198 nm and 202 nm when compared with
Purified water as reference liquid.Â
6.4.2Â
Acceptance criteria:Â Absorbance should be greater than 2 and transmittance
should be less than 1.0% T.
6.4.3Â Record
the calibration data on the SOP/FR-02Â Calibration datasheet.(Appendix II)
6.5 RESOLUTION: (for qualitative analysis)
6.5.1Â
0.020 % v/v solution of toluene in Hexane:Â Â
 Dilute
2ml of Toluene to 200ml with Hexane. Further, dilute 2 ml of the above solution to
100 ml with Hexane. Â
(Note:
Use only spectroscopy grade Toluene and Hexane)
6.5.2.
Record the spectrum in the range 260 to 275 nm of 0.020 % v/v toluene in Hexane
using Hexane in the reference cell.
6.5.3Â
Calculate the ratio of the absorbance at the maximum at about 269 nm to that at
the minimum at about 266 nm.
6.5.4
Acceptance criteria: The ratio of max/min should not be less than 1.5.
6.5.5Â Record
the calibration data on the SOP/FR-02Â Calibration datasheet. (Appendix
II).
6.6 RESOLUTION POWER (SECOND DERIVATIVE):
6.6.1
0.020 % v/v solution of toluene in methanol: Dilute 2 ml of Toluene to
200 ml with Methanol Further dilute 2 ml of the above solution to 100 ml with
Methanol. Â
 (Note: Use only spectroscopy grade Toluene and
methanol)
6.6.2Â
Record the second derivative spectrum in the range 255 to 275 nm of 0.020 %
v/v  toluene in methanol using methanol in the reference
6.6.3
Acceptance criteria:Â
A small negative extrema located between two large negative extrema at about 261
and 268 nm should be clearly visible. The ratio of max/min should not be less than 0.2
6.6.4
Record the calibration data on the SOP/FR-02 Calibration datasheet. (Appendix II)
6.7 SPECIFICATION OF CELLS:
6.7.1
In Photometry mode select transmittance mode. Set the wavelength at 200 nm.
Then press auto-zero to get 100 % transmittance with air blank in both the cell
holders. Then place the cell in the sample compartment filled with purified
water using air blank and read the transmittance. Similarly, repeat the
operation for 220 nm and 240 nm. Repeat the test for the second cell. The
transmittance values should meet the requirement as given in the following
table.
6.7.2 Acceptance criteria:
               Wavelength                                      %
Transmittance
               200 nm                                    Â
   Not Less Than 73%
               220 nm                                                Not Less Than 85%
               240 nm                                    Â
  Not Less Than 87%
               365 nm                                                Not Less Than 90%
               Note: Transmittance of one cell
is within ± 1.5% of the other cell.
6.7.3
Record the calibration data on APL/CQA/SOP-042/FR-02 Calibration datasheet.
(Appendix II)
6.8 I0 - LINE FLATNESS Test
6.8.1
With an air blank in the sample and reference compartment, perform the test.           Â
6.8.2Â Â Scan in the wavelength range 220 to 700 nm.
6.8.3
Obtain the spectrum in the % transmittance mode.
6.8.4
Acceptance criteria: The Io line obtained should not deviate from a horizontal
line at any point by more than 99.54 - 100.46 % transmittance.              Â
6.8.5
Record the Calibration Data on SOP/FR-02Â Calibration Data Sheet (Appendix
II)
6.8.6
Summarize the calibration data on SOP/FR-03 UV-VIS Summary sheet (Appendix
III)Â
7.0 GENERATION OF INSTRUMENT CALIBRATION NUMBER
Generate the Instrument Calibration number on the calibration datasheet as INSCALXXYYZZZ Where INS
denotes Instrument, CAL denotes Calibration XX denotes year; YY denotes Month
ZZZ denotes sequence number.
 8.0 ABBREVIATIONS
INS CAL:
Instrument Calibration Â
IP
& EPÂ Â :Â Indian Pharmacopoeia & European
PharmacopoeiaÂ