SOP for Operation and Calibration of Turbidity Meter | SOP for Calibration of Turbidity Meter

What:- This document details the procedure for the Operation and Calibration of Turbidity Meter CL-52D (Naphalometer) - Elico.

Why:- It is the policy of Xyz Pharmaceuticals Limited that a written procedure shall be followed for the Operation and Calibration of Turbidity Meter CL-52D (Naphalometer) - Elico and its use monitored to ensure smooth operation and its use monitored to obtain consistent and reproducible results and to minimize downtime and to meet regulatory requirements.

When:- This procedure is to be applied at the time of Operation and Calibration of Turbidity Meter CL-52D (Naphalometer) - Elico.

RESPONSIBILITY 

Persons along with their responsibilities and accountability are given below:

To prepare the SOP. To maintain the records as per SOP

To follow the SOP accordingly

Operation and Calibration of Turbidity Meter

5.1 PRELIMINARY CHECK

5.1.1 Check if the instrument is clean and free from dust, if not; clean it with a soft cloth duster.

5.1.2 Ensure that the temperature is about ambient and relative humidity is about 45%.

5.2 PRECAUTIONS

5.2.1 Allow the turbid meter to warm up for at least 15 minutes prior to use.

5.2.2  Always use clean glassware and handle cuvettes so there are no fingerprints in the area where light passes through the sample.

5.2.3 Use alcohol and a lint-free cloth to clean the outside of the cuvettes.

5.2.4 Once the cuvette becomes scratched or pitted, replace it with a new cuvette.

5.2.5 Rinse the cuvette with about 0.5 ml of the standard to be used. Discard. Repeat with a second rinse to ensure that only the standard is present in the cuvette.

5.2.6 Store the stock solution in a refrigerator at 4º C

5.2.7 Do not pour the used primary standard back into the bottle. This contaminates the contents.

5.2.8 Allow the standard to reach room temperature before using it to alleviate fogging of the cuvette.

5.2.9  If problems are encountered with taking a reading, allow the sample to sit loosely capped with a clean cap until all air bubbles have settled out or gently tap the sides of the cuvette to release air bubbles, and then take another reading.

5.3 OPERATION

5.3.1 Connect the plug to the main power supply and to the main switch of the instrument. Allow the turbid meter to warm up for at least 15 minutes.

5.3.2 To select the appropriate range Press the button push-out condition for 0-100 NTU and push-in condition for 100-1000 NTU given below side of the instrument.

5.3.3 Put the blank (Milli Q water filtered through 0.45µm membrane) into the sample holder and set the zero by zero setting knob.

5.3.4 Put the 100NTU turbidity solution into the sample holder and set 100NTU by the 100 setting knob.

5.3.5 Now put the unknown sample into the sample holder and note the reading.

5.3.6 Switch off the instrument when not in use.

5.3.7 Maintain the Logbook of Turbidity Meter on XYZ/CQA/SOP-074/FR-01 Logbook for Turbidity Meter Appendix I.

6.0 CALIBRATION PROCEDURE

Turbidity meter calibration solution

6.0.2.1 Preparation of stock standard turbidity solutions:

Solution (1):

Take 1.0gm of Hydrazine sulfate into a 100 ml volumetric flask and make up the volume with turbid-free water.

Operation and Calibration of Turbidity Meter

Solution (2):

Take 10.0 gm of Hexamethylenetetramine into the 100 ml volumetric flask and make up the volume with turbid-free water.

Stock 400 NTU solutions (3):

Pipette out 5.0ml stock solution (1) and 5.0ml stock solution (2) into the 100ml volumetric flask and stand for 24 hr at 25±2º C then fill up to the mark.

6.0.2.2 Preparation of intermediate standard solution:

40 NTU solutions (4):

Pipette 10.0ml of solution (4) into 100 ml volumetric flask and make up the volume up to the mark with turbid-free water.

6.0.2.3 Preparation of working standard solutions:

2 NTU Solution:

Pipette 5.0ml of solution (4) into 100 ml volumetric flask and make up the volume up to the mark with turbid-free water.

4 NTU Solution:

Pipette 10.0ml of solution (4) into a 100 ml volumetric flask and make up the volume to the mark with turbid-free water.

6 NTU Solution:

Pipette 15.0ml of solution (4) into a 100ml volumetric flask and make up the volume to the mark with turbid-free water.

8 NTU Solution:

Pipette 20.0ml of solution (4) into a 100 ml volumetric flask and make up the volume to the mark with turbid-free water.

10 NTU Solution:

Pipette 25.0ml of solution (4) into a 100 ml volumetric flask and make up the volume to the mark with turbid-free water.

Acceptance criteria

The linearity coefficient ‘r’ obtained from the linearity graph of Hydrazine sulfate and Hexamethylenetetramine for different levels should not be less than 0.999.

6.0.2.2 Discard the working turbidity solution after the calibration/verification.

6.0.2.3 Record the calibration data as per XYZ/CQA/SOP-074/FR-02 Appendix II 

6.0.3 DAILY VERIFICATION OF TURBIDITY METER:

6.0.3.1 For the verification, determine the turbidity of the following standard solutions by following the same procedure as specified under basic operation. The verification is to be done on a daily basis before using the instrument.

6.0.3.2 Record the verification data as per XYZ/CQA/SOP-074/FR-03, Appendix III 

6.0.3.3 Report any discrepancy observed during the operation and calibration of the instrument to Section

In charge of his representative for corrective and preventative action as per SOP No. XYZ/CQA/SOP-030, CAPA.

6.0.3.4 Section In-charge or his representative will take the necessary action and report the same to Manager-Quality Assurance.

6.0.3.5 Affix the “Out of Order” label on the instrument as per SOP No. XYZ/CQA/SOP-006, Labelling Policy.

7.0 GENERATION OF INSTRUMENT CALIBRATION NUMBER    

Generate the instrument calibration number on the calibration datasheet as INSCALXXYYZZZ where INS denotes instrument, CAL denotes calibration, XX denotes the year, YY denotes month and ZZZ denotes sequence number.

8.0 ABBREVIATION

XYZ               Xyz Pharmaceuticals Limited

SOP                 Standard Operating Procedure

QC                   Quality Control

CQA                Corporate Quality Assurance

LogBook and Appendix:- Download

>SOP for Operation and Calibration of Micropipettes

BANTI SINGH

Hi I'm Banti Singh, a Chemical Engineer! Welcome all of you to my blog. If you got the information right? Share the information. All of you Thank you

Thanks to visit this site.

Post a Comment (0)
Previous Post Next Post